      • 霍尼韋爾氣體傳感器MIDAS-E-NH3
        發布者:無錫遠路貿易有限公司  發布時間:2021-02-19 13:29:20  訪問次數:708

        Midas Cartridge Detectable Gases
        Gas Name Chemical Formula Range Sensor Part Number
        Ammonia NH3 9-100 ppm MIDAS-E-NH3
        Arsine AsH3 18-200 ppb MIDAS-E-ASH
        Boron Trichloride BCl3 0.72-8 ppm MIDAS-E-HCL
        Boron Trifluoride BF3 0.72-8 ppm MIDAS-E-HFX
        Boron Trifluoride (Low Level) BF3 0.18-2 ppm MIDAS-E-HFL
        Bromine Br2 0.036-0.4 ppm MIDAS-E-BR2
        Carbon Dioxide CO2 0.15-2.0% MIDAS-E-CO2
        Carbon Monoxide CO 9-100 ppm MIDAS-E-COX
        Chlorine Cl2 0.18-2 ppm MIDAS-E-HAL
        Chlorine Dioxide ClO2 0.036-0.4 ppm MIDAS-E-BR2
        Diborane B2
        H6 36-400 ppb MIDAS-E-B2H
        Dichlorosilane H2
        Cl2
        Si 0.72-8 ppm MIDAS-E-HCL
        Difluoromethane** CH2
        F2 16-240 ppm MIDAS-E-XCF
        Disilane Si2
        H6 1.8-20 ppm MIDAS-E-SHX
        Fluorine F2 0.36-4 ppm MIDAS-E-HAL
        Germane GeH4 70-800 ppb MIDAS-E-ASH
        Hexafluorobutadiene** C4
        F6 3-40 ppm MIDAS-E-CFX
        Hydrogen (%LEL) H2 6.5-100% LEL MIDAS-E-LEL*
        Hydrogen (ppm) H2 90-1000 ppm MIDAS-E-H2X
        Hydrogen Bromide HBr 0.72-8 ppm MIDAS-E-HCL
        Hydrogen Chloride HCl 0.72-8 ppm MIDAS-E-HCL
        Hydrogen Cyanide HCN 1.8-20 ppm MIDAS-E-HCN
        Hydrogen Fluoride HF 1.05-12 ppm MIDAS-E-HFX
        Hydrogen Fluoride (Low Level) HFL 0.18-2 ppm MIDAS-E-HFL
        Hydrogen Sulfide H2
        S 3.6-40 ppm MIDAS-E-H2S
        Methane (%LEL) CH4 6.5-100% LEL MIDAS-E-LEL*
        Methyl Fluoride** CH3
        F 8-120 ppm MIDAS-E-XHF
        Nitric Oxide NO 9-100 ppm MIDAS-E-NOX
        Nitrogen Dioxide NO2 1.05-12 ppm MIDAS-E-NO2
        Nitrogen Trifluoride** NF3 3.6-40 ppm MIDAS-E-HFX for 00P, XHF for NP1
        Octofluorocyclopentene** C5
        F8 3-40 ppm MIDAS-E-XCF
        Oxygen O2 0.2-25% v/v MIDAS-E-O2X
        Ozone O3 0.065-0.7 ppm MIDAS-E-O3H
        Ozone (Low Level) O3 0.036-0.4 ppm MIDAS-E-O3X
        Phosphine PH3 110-1200 ppb MIDAS-E-PH3
        Silane SiH4 1.8-20 ppm MIDAS-E-SHX
        Silane (Low Level) SiH4 0.18-2 ppm MIDAS-E-SHL
        Sulfur Dioxide SO2 0.7-8 ppm MIDAS-E-SO2
        Tetra Ethyl Ortho Silicate TEOS 3.6-40 ppm MIDAS-E-TEO
        Tungsten Hexafluoride WF6 1.05-12 ppm MIDAS-E-HFX
        Tungsten Hexafluoride (Low Level) WF6 0.18-2 ppm MIDAS-E-HFL

        Midas H2氫氣氣體探測傳感器MIDAS-S-H2X MIDAS-E-H2X

        Hydrogen ppm (H2)
        MIDAS-S-H2X
        MIDAS-E-H2X
        Midas? sensor cartridges are intended
        for use only with Honeywell Analytics’
        Midas? Gas Detector System. Please
        refer to the Midas? Technical Manual
        for further details.
        Technical Data
        Gas Measured Hydrogen (H2)
        Cartridge Part Number MIDAS-S-H2X 1 year standard warranty
        MIDAS-E-H2X 2 year extended warranty
        Sensor Technology 3 electrode electrochemical cell
        Measuring range (ppm) H2 0 – 1000 ppm
        Minimum Alarm 1 Set Point 120 ppm
        Lower Detectable Limit (LDL) 90 ppm
        Repeatability < ? 5 % of measured value
        Linearity < ? 5 % of measured value
        Response Time t62.5 < 50 seconds
        Sensor Cartridge Life Expectancy ≥ 24 months under typical application conditions
        Operating Temperature 0? to 40?C (32? to 104?F)
        Effect of Temperature < ? 0.17 ppm / ?C (0? to 20?C )
        Zero < ? 0.07 ppm / ?C (20? to 40?C )
        Sensitivity < ? 3 % of measured value / ?C
        Operating Humidity (continuous) 10 – 90 % rH
        Effect of Humidity
        Zero No Effect
        Sensitivity No Effect
        Operating Pressure 90 – 110 kPa
        Effect of bbbbbbbb No effect in typical application
        Long Term Drift
        Zero No Drift
        Sensitivity < ? 10 % of measured value / 6 month
        Calibration Gas Hydrogen (H2)
        Challenge Gas (Bump Test) Hydrogen (H2)
        Warm Up time < 10 minutes
        Storage Temperature 5? to 25?C ( 41? to 77?F)
        The sensor data listed is based on ideal test environment; observed performance may vary based on the actual monitoring system and the sampling conditions employed
        Gas / Vapor Chemical Formula Concentration applied (ppm) Reading (ppm H2)
        Ammonia NH3 100 0
        Arsine AsH3 0.2 0
        Carbon Dioxide CO2 1000 0
        Carbon Monoxide CO 100 150
        Chlorine Cl2 1 0
        Chlorine Dioxide ClO2 1 0
        Hydrogen Cyanide HCN 20 0
        Hydrogen Sulfide H2S 20 4
        Iso Propanol C3H7OH 1100 Yes
        Methane CH4 1% 0
        Nitrogen Dioxide NO2 10 -40
        Ozone O3 0.25 0
        Sulfur Dioxide SO2 5 0
        MDA Scientific has developed a sophisticated range of highly sensitive gas detection equipment, designed to perform in ways that define new gas detection performance levels providing total solutions to protect people, improve production efficiency and reduce costs.

        Midas? is a registered trademark of Honeywell Analytics.
        As with all electrochemical sensor cells, dramatic output changes in reported concentrations can be
        expected under rapidly changing environmental conditions. Please ensure sensors are located in areas not
        prone to sudden changes in humidity and temperature.
        Actual readings may be affected by flow rates (although Midas? automatically controls flow rate within
        specified ranges) and absorption on tubing and other gas path surfaces.
        All sensors are shipped pre-calibrated to traceable national standards. Dependent on actual operating
        conditions and overall exposure to gases, each sensor may not require in field calibration for up to 24
        months subject to any requirements to calibrate from local regulations or site practices.
        Calibration and challenge gases should be from a certified and reliable source.
        Cross Sensitivities
        Each Midas? sensor is potentially cross sensitive to other gases and this may cause a gas reading when
        exposed to other gases than those originally designated. The table below presents typical readings that
        will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases
        containing the cross sensitive species).